Clean Hood

Made in the UK
Our products are proudly designed and manufactured in the UK

Flexible booths, simply made to suit your process

The Clean Hood laminar flow hood (sometimes called a laminar flow cabinet) provides a benchtop, particle-free working environment for your process. The design allows you to add features as your process and budget require.

The Clean Hood design is based on the (FFU) by Moorfield Nanotechnology, which is the backbone of the successful Clean Tent range of products.

Clean Hood features:

  • ISO 5/ Class 100
  • Anodised aluminium frame
  • 6mm clear acrylic sheet to 3 sides, designed to allow as much ambient light into the work station as possible
  • Variable speed fan filter unit controller
  • Vertical and Horizontal flow available on all models



The Clean Hood has been designed with longevity in mind.  The variable speed controller enables you to set the face velocity, ensuring that you do not over pressurise the HEPA filter.

Unlike traditional enclosed units both the pre-filter and the final H14 HEPA filter are easy to replace.

 

Get a quote

Complete a short form and and a member of the Clean Environments team will send you a quotation*

*In some cases, additional information may be requested.

Or ask an expert about Clean Hood















Clean Hood Models

Vertical Laminar Flow Hood

Clean Hood CH1800V laminar flow hood

CH1800V

Footprint 1812mm(L) x 612mm(D) x 1163mm(H)
Workspace 1780mm(L) x 590mm(W) x 750mm(H)
Footprint (extra deep option) 1812mm(L) x 912mm(D) x 1163mm(H)
Workspace (extra deep option) 1780mm(L) x 890mm(W) x 750mm(H)
Request a quote

 

Clean Hood CH1200V laminar flow hood

CH1200V

Footprint 1212mm(L) x 612mm(D) x 1163mm(H)
Workspace 1180mm(L) x 590mm(W) x 750mm(H)
Footprint (extra deep option) 1212mm(L) x 912mm(D) x 1163mm(H)
Workspace (extra deep option) 1180mm(L) x 890mm(W) x 750mm(H)
Request a quote

Horizontal Laminar Flow Hood

CH1800H

Footprint 1812mm(L) x 1013mm(D) x 800mm(H)
Workspace 1780mm(L) x 590mm(W) x 750mm(H)
Footprint (extra deep option) 1812mm(L) x 1313mm(D) x 800mm(H)
Workspace (extra deep option) 1780mm(L) x 890mm(W) x 750mm(H)
Request a quote

 

CH1200H

Footprint 1212mm(L) x 1013mm(D) x 800mm(H)
Workspace 1180mm(L) x 590mm(W) x 750mm(H)
Footprint (extra deep option) 1212mm(L) x 1313mm(D) x 800mm(H)
Workspace (extra deep option) 1180mm(L) x 890mm(W) x 750mm(H)
Request a quote

 

Clean Hood Case Studies

View All Clean Hood Case Studies

Clean Hood Specifications

All Clean Hoods are supplied with the following components. For more information, please contact us

Electrical

Filter unit fan

280W, 230vac 50Hz standard

Electrical Supply

5A, 230vac 50Hz single phase 3 pin plug.

Fan Filter Unit

Lifespan

Full speed 20,000 hrs continuous

HEPA H14

Vmax at face

0.45 m/sec.

Expected HEPA lifespan

5 years in lab conditions with continuous fan

Pre-filter

G4 pleated pre filter

475mm x 370mm.

Expected filter lifespan

Approximately 3 months in lab conditions with continuous fan.

 